RDP Home Page Banner

Model FP2000 Pressure Transducer


Very high accuracy
Internal amplifier
Stainless steel
Absolute pressure
Gauge pressure
Differential pressure
High resolution
image of Model  FP2000 Pressure Transducer
The FP2000 series offers a modular selection which increases choice and reduces delivery time. We try to stock the various elements and simply weld them together to your requirements.
This unit is available in gauge, absolute, differential and barometric versions..
The SPT-FP2000 gauge pressure transducer measures with respect to atmospheric pressure.
The SPT-FP2000 absolute pressure transducer measures with respect to a vacuum.
A wet/wet differential pressure transducer measures the difference in the pressure applied to its two ports, both ports may contain liquid or gas.
This results in a wide range of mechanical and electrical options and in a shorter lead-time.

Type
FPG Gauge pressure The SPT-FP2000 gauge pressure transducer measures with respect to atmospheric pressure
FPA Absolute pressure The SPT-FP2000 absolute pressure transducer measures with respect to a vacuum
FPB Barometric pressure A barometric transducer measures low pressure levels with respect to a vacuum
FPV Vacuum pressure A vacuum transducer measures the depression below atmospheric pressure
FDD Differential pressure (Wet/dry) A wet/dry differential pressure transducer measures the difference in the pressure applied to its two ports, one of the ports can only contain a dry gas
FDW Differential pressure (Wet/wet) A wet/wet differential pressure transducer measures the difference in the pressure applied to its two ports, both ports may contain liquid or gas
FPA, FPB, FPG, FPV FDW FDD
dimensional drawing of  Model  FP2000 Pressure Transducer image of FDW image of FDD
D = 29mm
L1= 74mm (≤5psi)
L1= 81mm (≥15psi)
D = 29mm
L1 = 97mm
L2 = 38mm
D = 29mm
L1 = 104mm
L2 = 23mm

Accuracy
1 0.1% Thermal error band = 0.5%. (Not available with output 2u (unamplified))
2 0.25% Thermal error band = 1%

Range
Cable Option Range Types allowed
UQ 16 to 32''Hg (54 to 108kPaA) FPB
UR 26 to 32''Hg (88 to 108kPaA) FPB
UG 0 to 30''Hg (0 to 102kPaA) FPB
WA 10''H2O (2.5kPa) FPG, FPV, FDD & FDW
AP 1psi (7kPa) FPG, FPV, FDD & FDW
AT 5psi (34kPa) FPG, FPA, FPV, FDD & FDW
AV 10psi (69kPa) FPG, FPA, FPV, FDD & FDW
BJ 15psi (103kPa) FPG, FPA, FPV, FDD & FDW
BM 30psi (207kPa) FPG, FPA, FDD & FDW
BN 50psi (345kPa) FPG, FPA, FDD & FDW
BR 100psi (689kPa) FPG, FPA, FDD & FDW
CJ 150psi (1MPa) FPG, FPA, FDD & FDW
CN 250psi (2MPa) FPG, FPA, FDD & FDW
CR 500psi (3MPa) FPG, FPA, FDD & FDW
CT 750psi (5MPa) FPG, FPA, FDD & FDW
CV 1000psi (7MPa) FPG, FPA, FDD & FDW
DJ 1500psi (10MPa) FPG & FPA
DM 2500psi (17MPa) FPG & FPA
DR 5000psi (34MPa) FPG & FPA
DT 7500psi (52MPa) FPG & FPA
DV 10000psi (69MPa) FPG & FPA

Output
2u mV/V Supply voltage = 0 to 10V. Sensitivity (nom) 2.5 to 10mV/V
2e 0 to 5V Supply voltage = 9 to 28V
2f 0 to 10V Supply voltage = 15 to 28V
2n 4-20mA (Supply Current ,≤5000psi) Supply voltage = 9 to 28V
2y 4-20mA Supply voltage = 9 to 32V

Pressure port thread
Cable Option 5a Cable Option 5b Cable Option 5c Cable Option 5d Cable Option 5f Cable Option 5g
1/4-18 NPT female
H= 13mm (≤1000psi)
H= 25mm (≥1500psi)
1/4-18 NPT male
H= 28mm (≤1000psi)
H= 33mm (≥1500psi)
7/16-20 UNF female
H= 15mm (≤1000psi)
H= 25mm (≥1500psi)
7/16-20 UNF male
H= 25mm (≤1000psi)
H= 33mm (≥1500psi)
G1/4B female
H= 18mm (≤1000psi)
H= 25mm (≥1500psi)
G1/4B male
H= 30mm (≤1000psi)
H= 30mm (≥1500psi)
Option 5a Drawing dimensional drawing of Cable Option 5b dimensional drawing of Cable Option 5c dimensional drawing of Cable Option 5d dimensional drawing of Cable Option 5f dimensional drawing of Cable Option 5g

Electrical termination
Cable Option 6a Cable Option 6m Cable Option 6n Cable Option 6q Cable Option 6r
PT1H-10-6P (vented)
T = 7.6mm
DIN43650 (vented)
T = 5.1mm
DIN (vented)
T = 7.6mm
integral cable (vented)
Cable length = 1.5m
T = 7.6mm
integral cable with conduit fitting (vented)
Cable length = 1.5m
T = 7.6mm
dimensional drawing of Cable Option 6a dimensional drawing of Cable Option 6m dimensional drawing of Cable Option 6m dimensional drawing of Cable Option 6q dimensional drawing of Cable Option 6q

Compensated temperature range
Type Range Compensated temperature range Compensated temperature range (Cable Option 1y)
Thermal error band = ±1%
FPG & FPA 10''H2O (2.5kPa) 4℃ to 43℃ -18℃ to 82℃
1psi (7kPa) 4℃ to 43℃ -18℃ to 82℃
5psi (34kPa),10psi (69kPa) & 15psi (103kPa) 4℃ to 60℃ -18℃ to 82℃
>15psi (103kPa) 4℃ to 60℃ -18℃ to 82℃
FPB 26 to 32''Hg
(88 to 108kPaA)
16℃ to 43℃ -1℃ to 77℃
16 to 32''Hg
(54 to 108kPaA)
4℃ to 60℃ -1℃ to 77℃
0 to 30''Hg
(0 to 102kPaA)
4℃ to 43℃ -1℃ to 77℃
FPV 10''H2O (2.5kPa) 4℃ to 43℃ -18℃ to 77℃
1psi (7kPa) 4℃ to 43℃ -18℃ to 77℃
5psi (34kPa),10psi (69kPa) & 15psi (103kPa) 4℃ to 60℃ -18℃ to 77℃
FDD & FDW 10''H2O (2.5kPa) & 5psi (34kPa) 4℃ to 43℃ -20℃ to 80℃
>5psi (34kPa) 4℃ to 60℃ -20℃ to 80℃

General specification
  FPG/FPA FPV FPB FDW FDD
Over-range capacity ≤1000psi 4X F.S. (3000psi maximum)
≥15000psi 4X F.S. (15000psi maximum)
≤1000psi 4X F.S. (3000psi maximum) ≤1000psi 4X F.S. (3000psi maximum) Port 1 = 4X F.S. (3000psi maximum)
Port 2 = 4X F.S. (250psi maximum)
Port 1 = 4X F.S. (3000psi maximum)
Port 2 = 4X F.S. (250psi maximum)
Line pressure - - - 500 psi 500p psi
Port material Ha C276 & 316L ss Ha C276 & 316L ss Ha C276 & 316L ss Ha C276 & 316L ss Port 1 = Ha C276 & 316L ss
Port 2 = Dry air

Options and accessories
Amplifiers Amplifier option thumbnail
Panel meter Panel meter option thumbnail
3d Remote buffered shunt cal
14c Side access to adjustment potentiometers  ( +/- 10% )

Example - FDW-1-CN-2y-5b-6a-1y-0-0
Type Differential pressure (Wet/wet) FDW
Accuracy 0.1% 1
Range 250psi (2MPa) CN
Output 4-20mA 2y
Pressure port thread 1/4-18 NPT male 5b
Electrical termination PT1H-10-6P (vented) 6a
Cable Option 1y Compensated temperature range (Cable Option 1y) 1y
Cable Option 3d - 0
Cable Option 14c - 0