LVDT amplifier
Strain gauge transducer amplifier
Digital display module
Limit trips
Serial output
Voltage / 4-20mA output
Simple trimpot controls
Signal conditioning is required where the output of a transducer needs to be boosted or changed into a form suitable for the monitor or logging device which will be used. A very wide range of gain adjustment ensures that our amplifiers are compatible with the vast majority of LVDT and strain gauge sensors available from any manufacturer.
Modular 600 signal conditioning allows the customer to select the type and quantity of input cards required for the application from a number of different types available. The system can be supplied with optional monitor/display/data logger units if required.
Firstly decide the type and quantity of modules required making sure to include any future requirements. Then decide if you need a monitor or data logger. Finally choose an appropriately sized housing with a power supply and blank panels to cover any unused slots.
Modular 600 signal conditioning allows the customer to select the type and quantity of input cards required for the application from a number of different types available. The system can be supplied with optional monitor/display/data logger units if required.
Firstly decide the type and quantity of modules required making sure to include any future requirements. Then decide if you need a monitor or data logger. Finally choose an appropriately sized housing with a power supply and blank panels to cover any unused slots.
611 Strain gauge transducer amplifier
Compatible with most full bridge strain gauge transducers. Most amplified transducers from any manufacturer.
Number of channels | 2 |
---|---|
Supply voltage (dual) | ±15V, 60mA (plus transducer and output load). |
Output 1 | ±10V |
Output 2 | 4 to 20mA (loop resistance 0Ω to 450Ω) |
Output ripple | 4mV/15uA peak-to-peak (typical) |
Approximate zero adjustment range | ±10V |
Analogue output bandwidth | 200/10Hz |
Signal input range | -13V to +13V |
Amplifier gain range | 1 to 2000 |
Transducer excitation | 1V to 10V, 110mA maximum |
Linearity error | ±0.02% F.S. (typical) |
Temperature coefficient (zero) | ±0.001% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.002% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
615 RDP PY series amplifier
Compatible with series PY Extreme environment non-contact displacement transducer
Number of channels | 1 |
---|---|
Supply voltage (dual) | ±15V, 60mA (typical) |
Output 1 | ±10V |
Output 2 | 4 to 20mA (loop resistance 0Ω to 450Ω) |
Output ripple | 10mv/120uA peak-to-peak (typical) |
Approximate zero adjustment range | ±10V |
Analogue output bandwidth | 160/20Hz |
Transducer excitation | 2.4kHz/4.8kHz/9.6kHz 10mA, (17Vrms maximum) |
Linearity error | ±0.05% F.S. (typical) |
Temperature coefficient (zero) | ±0.003% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.011% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
621 LVDT amplifier
Compatible with any standard RDP LVDT (without integral electronics). Most LVDTs from any manufacturer.
Number of channels | 2 |
---|---|
Supply voltage (dual) | ±15V, 65mA (typical) |
Output 1 | ±10V |
Output 2 | 4 to 20mA (loop resistance 0Ω to 450Ω) |
Output ripple | 10mv/100uA peak-to-peak (typical) |
Approximate zero adjustment range | 10V |
Analogue output bandwidth | 400Hz |
Signal input range | 0.012V to 4V |
Amplifier gain range | 2.5 to 833 |
Transducer excitation | 1.1V, 5kHz rms, 100mA maximum |
Linearity error | ±0.05% F.S. (typical) |
Temperature coefficient (zero) | ±0.003% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.003% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
626 LVDT amplifier with Sum/Average/Difference calculation
Any standard RDP LVDT (without integral electronics). Most LVDTs from any manufacturer.
Number of channels | 2 |
---|---|
Supply voltage (dual) | ±15V, 65mA (typical) |
Output ripple | 10mv/100uA peak-to-peak (typical) |
Approximate zero adjustment range | ±10V |
Analogue output bandwidth | 500Hz |
Signal input range | 0.12V to 4V |
Amplifier gain range | 2.5 to 833 |
Transducer excitation | 1.1V, 5kHz rms, 100mA maximum |
Linearity error | ±0.05% F.S. (typical) |
Temperature coefficient (zero) | ±0.003% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.003% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
628 1/4, 1/2 and full bridge strain gauge amplifier
Compatible with 1/4, 1/2 and full bridge strain gauges. Most full bridge strain gauge transducers.
Number of channels | 1 |
---|---|
Supply voltage (dual) | ±15V, 70mA (plus transducer and output load) |
Output 1 | ±10V |
Output 2 | 4 to 20mA (loop resistance 0Ω to 450Ω) |
Output ripple | 40mv/80uA peak-to-peak (typical) |
Approximate zero adjustment range | ±0.4V |
Analogue output bandwidth | 10/100/1k/10kHz |
Signal input range | -11V to +11V |
Amplifier gain range | 1 to 80000 |
Transducer excitation | 1V to 10V, 110mA maximum |
Linearity error | ±0.02% F.S. (typical) |
Temperature coefficient (zero) | ±0.0006% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.001% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
681 Limit trip module
Compatible with the voltage output of any modular 600 amplifier.
Number of channels | 4 |
---|---|
Supply voltage (dual) | ±15V, 100mA |
Signal input range | ±10V |
Limit trips:-Operation time | 6ms |
Limit trips:-Contact rating (maximum) | 30Vdc,1A /125Vac, 60VA |
Accuracy | ±0.01% F.S. (typical) |
Operating temperature range | 32℉ to 140℉ |
635 Digital display module
Number of channels | 24 (maximum) |
---|---|
Number of digits | 4½, 10mm LED |
Accuracy | ±0.03% F.S. (typical) |
Linearity error | ±0.05% F.S. (typical) |
Temperature coefficient (zero) | ±0.001% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.003% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
650 Serial interface and data logger module
Number of channels | 164 (maximum) |
---|---|
Number of digits | 4½, 10mm LED |
Baud rate | 600 to 57600 |
Accuracy | ±0.03% F.S. (typical) |
Temperature coefficient (zero) | ±0.001% F.S. /℉ (typical) |
Temperature coefficient (span) | ±0.003% F.S. /℉ (typical) |
Operating temperature range | 32℉ to 140℉ |
631 Power supply
Supply voltage | 94 to 253Vac, 80W |
---|---|
Operating temperature range | 32℉ to 131℉ |
602 housing
A=10.6''
B=9.9''
C=9.3''
D=11.1''
E=9.8''
F=5.2''
G=2.2''
H=0.7''
J=11.0''
K=0.3''
603 housing
A=14.8''
B=14.1''
C=13.5''
D=11.1''
E=9.8''
F=5.2''
G=2.2''
H=0.7''
J=15.2''
K=0.3''
604 housing
A=19.0''
B=18.0''
C=17.7''
D=11.1''
E=9.8''
F=5.2''
G=2.2''
H=0.7''
J=19.4''
K=0.3''